Please use this identifier to cite or link to this item: https://olympias.lib.uoi.gr/jspui/handle/123456789/8692
Full metadata record
DC FieldValueLanguage
dc.contributor.authorYao, D. C.en
dc.contributor.authorVlessidis, A. G.en
dc.contributor.authorEvmiridis, N. P.en
dc.date.accessioned2015-11-24T16:43:23Z-
dc.date.available2015-11-24T16:43:23Z-
dc.identifier.issn0003-2670-
dc.identifier.urihttps://olympias.lib.uoi.gr/jspui/handle/123456789/8692-
dc.rightsDefault Licence-
dc.subjectchemiluminescenceen
dc.subjectnitric oxideen
dc.subjectfiber optic sensorsen
dc.subjectonline detectionen
dc.subjectin-vivoen
dc.subjectsynthaseen
dc.subjectassayen
dc.subjectmicrosensoren
dc.subjectelectrodesen
dc.subjectnoen
dc.titleOn-line monitoring of nitric oxide complexed with porphyrine-bearing biochemical materials by using flow injection with chemiluminescence detectionen
heal.typejournalArticle-
heal.type.enJournal articleen
heal.type.elΆρθρο Περιοδικούel
heal.identifier.primaryDoi 10.1016/S0003-2670(01)00875-3-
heal.identifier.secondary<Go to ISI>://000168911800007-
heal.identifier.secondaryhttp://ac.els-cdn.com/S0003267001008753/1-s2.0-S0003267001008753-main.pdf?_tid=8bf09be0-356a-11e3-93a7-00000aacb361&acdnat=1381821928_ab998bd96cabf8eda6d650b7f4ae38f6-
heal.languageen-
heal.accesscampus-
heal.recordProviderΠανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Χημείαςel
heal.publicationDate2001-
heal.abstractAn on-line nitric oxide (NO) monitoring method is described employing a flow injection manifold with chemiluminescence (CL) detection. The method is tailored for monitoring the NO from its complexes with porphyrin ring-bearing (PRB) biochemical materials. The method is optimized for sensitivity and detection limit. The method is quite sensitive, simple, rapid, and precise. A sensitivity of 2.63 x 10(5) mV mol(-1) NO, a range of 2.0 x 10(-5) up to 1.0 x 10(-2) mol NO (under the conditions used) a reproducibility of 1.0% over the range 1.0 x 10(-3) up to 1.0 x 10-2 mol NO, a frequency of ca. five samples per minute and a detection limit of 1.0 x 10(-5) mol NO are the analytical figures of merit for the proposed method. (C) 2001 Elsevier Science B.V. All rights reserved.en
heal.publisherElsevier Massonen
heal.journalNameAnalytica Chimica Actaen
heal.journalTypepeer reviewed-
heal.fullTextAvailabilityTRUE-
Appears in Collections:Άρθρα σε επιστημονικά περιοδικά ( Ανοικτά). ΧΗΜ

Files in This Item:
File Description SizeFormat 
Yao-2001-On-line monitoring o.pdf151.75 kBAdobe PDFView/Open    Request a copy


This item is licensed under a Creative Commons License Creative Commons