dc.titleAssessment of the performances of three oscillometric blood pressure monitors for neonates using a simulatoren
heal.typejournalArticle-
heal.type.enJournal articleen
heal.type.elΆρθρο Περιοδικούel
heal.identifier.secondaryhttp://www.ncbi.nlm.nih.gov/pubmed/10362888-
heal.languageen-
heal.accesscampus-
heal.recordProviderΠανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών Υγείας. Τμήμα Ιατρικήςel
heal.publicationDate1999-
heal.abstractBACKGROUND: The majority of monitors for non-invasive measurement of blood pressure in neonates and infants use the oscillometric method. OBJECTIVE: To use a simulator to investigate the overall system accuracy for three oscillometric blood pressure monitors. The tested devices are DINAMAP 8100 (Critikon), SpaceLabs M90426 (SpaceLabs Medical)and the module HP M 1008B (Hewlett-Packard). METHODS: The blood pressure values obtained by the three devices were compared with those of the invasive reference. A blood pressure simulator was used for testing the performance of the three blood pressure monitors. RESULTS: The Dinamap and SpaceLabs readings are generally in good agreement with the invasive reference. In contrast, the HP M 1008B readings for the diastolic and mean arterial pressures are inaccurate (the mean errors are 21 and 15mmHg, respectively). CONCLUSION: These results contradict our previously published data obtained when a different simulator (Cufflink)was used. These findings are important, because blood pressure values that are too low can result in a different therapeutic approach being applied, for newborn babies have only a small range of mean arterial pressure for the autoregulation of the cerebral blood flow.en
heal.journalNameBlood Press Moniten