Please use this identifier to cite or link to this item: https://olympias.lib.uoi.gr/jspui/handle/123456789/15989
Full metadata record
DC FieldValueLanguage
dc.contributor.authorBenis, E. P.en
dc.contributor.authorZouros, T. J. M.en
dc.date.accessioned2015-11-24T18:27:07Z-
dc.date.available2015-11-24T18:27:07Z-
dc.identifier.issn0168-9002-
dc.identifier.urihttps://olympias.lib.uoi.gr/jspui/handle/123456789/15989-
dc.rightsDefault Licence-
dc.subjectelectron spectroscopyen
dc.subjecthemispherical spectrometersen
dc.subjectfringing field correctionsen
dc.subjectposition-sensitive detectorsen
dc.subjectescaen
dc.subjectloss spectrometeren
dc.subjectelectronen
dc.titleImproving the energy resolution of a hemispherical spectrograph using a paracentric entry at a non-zero potentialen
heal.typejournalArticle-
heal.type.enJournal articleen
heal.type.elΆρθρο Περιοδικούel
heal.identifier.primaryDoi 10.1016/S0168-9002(99)00954-7-
heal.identifier.secondary<Go to ISI>://000085077600021-
heal.identifier.secondaryhttp://ac.els-cdn.com/S0168900299009547/1-s2.0-S0168900299009547-main.pdf?_tid=8b3e4ce2-4de8-11e3-9a5d-00000aacb362&acdnat=1384514871_7bf18fc797e091a8a2cd7b657bb088a1-
heal.languageen-
heal.accesscampus-
heal.recordProviderΠανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών και Τεχνολογιών. Τμήμα Βιολογικών Εφαρμογών και Τεχνολογιώνel
heal.publicationDate2000-
heal.abstractWe demonstrate by ray tracing analysis that the focusing properties of a hemispherical spectrograph depend critically on the placement of the electron entrance position R(0) and the value of the potential V(0) at R(0). Improvements over conventional spectrographs using R(0) = (R) over bar, the mean radius and V(0) = 0 are predicted for particular combinations of R(0) and V(0). A spectrograph with (R) over bar = 101.6 mm and an entrance at R(0) = 82.55 mm using a zoom lens and a two-dimensional position sensitive detector (2D-PSD) with 40 mm multichannel plates and resistive anode encoder has been used to confirm the predicted qualitative behaviour. These results suggest that improvement in energy resolution may be attained without the use of fringing field correctors. (C) 2000 Elsevier Science B.V. All rights reserved.en
heal.journalNameNuclear Instruments & Methods in Physics Research Section a-Accelerators Spectrometers Detectors and Associated Equipmenten
heal.journalTypepeer reviewed-
heal.fullTextAvailabilityTRUE-
Appears in Collections:Άρθρα σε επιστημονικά περιοδικά ( Ανοικτά)

Files in This Item:
File Description SizeFormat 
Benis-2000-Improving the energy.pdf239.23 kBAdobe PDFView/Open    Request a copy


This item is licensed under a Creative Commons License Creative Commons