Please use this identifier to cite or link to this item:
https://olympias.lib.uoi.gr/jspui/handle/123456789/13893
Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Voltairas, P. A. | en |
dc.contributor.author | Fotiadis, D. I. | en |
dc.contributor.author | Massalas, C. V. | en |
dc.date.accessioned | 2015-11-24T17:33:45Z | - |
dc.date.available | 2015-11-24T17:33:45Z | - |
dc.identifier.issn | 0304-8853 | - |
dc.identifier.uri | https://olympias.lib.uoi.gr/jspui/handle/123456789/13893 | - |
dc.rights | Default Licence | - |
dc.subject | retinal detachment | en |
dc.subject | retina elasticity | en |
dc.subject | eye surgery | en |
dc.subject | ferrofluid | en |
dc.subject | modeling | en |
dc.subject | anatomy of the eye | en |
dc.subject | scleral buckle | en |
dc.subject | silicone | en |
dc.subject | tamponade | en |
dc.subject | cells | en |
dc.subject | eye | en |
dc.title | Elastic stability of silicone ferrofluid internal tamponade (SFIT) in retinal detachment surgery | en |
heal.type | journalArticle | - |
heal.type.en | Journal article | en |
heal.type.el | Άρθρο Περιοδικού | el |
heal.identifier.secondary | <Go to ISI>://000168288100040 | - |
heal.identifier.secondary | http://ac.els-cdn.com/S0304885300012658/1-s2.0-S0304885300012658-main.pdf?_tid=aff535cf6d825cf77d998de152aee867&acdnat=1339758816_3bf0b8c523043b5032a654f75eba75a2 | - |
heal.language | en | - |
heal.access | campus | - |
heal.recordProvider | Πανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Μηχανικών Επιστήμης Υλικών | el |
heal.publicationDate | 2001 | - |
heal.abstract | It has been argued that silicone ferrofluid internal tamponade (SFIT) can provide (360 degrees) tamponade of the retina in retinal detachment surgery. Provided that the produced SFIT is biocompatible, exact knowledge is needed of its elastic stability in the magnetic field produced by the semi-solid magnetic silicon band (MSB) used as a scleral buckle. We propose a quantitative, phenomenological model to estimate the critical magnetic field produced by the MSB that 'closes' retinal tears and results in the reattachment of the retina. The magnetic 'deformation' of SFIT is modeled in accordance with the deformation of a ferrofluid droplet in an external magnetic field. (C) 2001 Elsevier Science B.V. All rights reserved. | en |
heal.publisher | Elsevier | en |
heal.journalName | Journal of Magnetism and Magnetic Materials | en |
heal.journalType | peer reviewed | - |
heal.fullTextAvailability | TRUE | - |
Appears in Collections: | Άρθρα σε επιστημονικά περιοδικά ( Ανοικτά) |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Voltairas-2001-Elastic stability of.pdf | 305.23 kB | Adobe PDF | View/Open Request a copy |
This item is licensed under a Creative Commons License