dc.subjectradiochromic film dosimetryen
dc.subjectintravascular brachytherapyen
dc.subjectradiation-therapyen
dc.subjectsource trainsen
dc.subjectquality-assuranceen
dc.subjectcalibrationen
dc.subjectcanceren
dc.titlePatient and personnel dosimetry in endovascular radiotherapy with Sr-90/Y-90 sourcesen
heal.typejournalArticle-
heal.type.enJournal articleen
heal.type.elΆρθρο Περιοδικούel
heal.identifier.primaryDoi 10.1093/Rpd/Nch496-
heal.identifier.secondary<Go to ISI>://000230595300007-
heal.identifier.secondaryhttp://rpd.oxfordjournals.org/content/114/4/514.full.pdf-
heal.languageen-
heal.accesscampus-
heal.recordProviderΠανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών Υγείας. Τμήμα Ιατρικήςel
heal.publicationDate2005-
heal.abstractfEndovascular brachytherapy (EVBT) is an established treatment to reduce the probability of restenosis after a percutaneous coronary intervention. The purpose of this study was to assess (1) the manufacturer's stated dosimetric data for Sr-90/Y-90 source trains to be used in EVBT and (2) the procedure-related radiation burden. The radiation fields in water around six Sr-90/Y-90 source trains were studied using phantoms made of 'solid water' and MD55-2 radiochromic films. The water equivalence of the phantom material was tested by applying quantitative computed tomography. Thermoluminescence dosemeters were used to assess personal radiation burden and crosscheck the dose distribution along the source trains. Technical failure was observed in one source train and this train was excluded from analysis. The measured dose rate in water at 2 mm radial distance was on average 8% higher than the manufacture's stated value (range of measured to stated values 1.05-1.15). The dose rate decreased exponentially with radial distance between 2 and 6 mm. The dose rate in contact with the source viewing window of the delivery devices ranged between 0.5 and 7.5 mGy h(-1). Low-energy photons were the main contributors to personal dose.en
heal.journalNameRadiation Protection Dosimetryen
heal.journalTypepeer-reviewed-
heal.fullTextAvailabilityTRUE-
Appears in Collections:Άρθρα σε επιστημονικά περιοδικά ( Ανοικτά) - ΙΑΤ

Files in This Item:
File Description SizeFormat 
Kalef-Ezra-2005-Patient and personne.pdf360.6 kBAdobe PDFView/Open    Request a copy


This item is licensed under a Creative Commons License Creative Commons