Please use this identifier to cite or link to this item:
https://olympias.lib.uoi.gr/jspui/handle/123456789/19439
Full metadata record
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Kalef-Ezra, J. | en |
dc.contributor.author | Bozios, G. | en |
dc.contributor.author | Tsekeris, P. | en |
dc.contributor.author | Michalis, L. | en |
dc.date.accessioned | 2015-11-24T18:59:54Z | - |
dc.date.available | 2015-11-24T18:59:54Z | - |
dc.identifier.issn | 0144-8420 | - |
dc.identifier.uri | https://olympias.lib.uoi.gr/jspui/handle/123456789/19439 | - |
dc.rights | Default Licence | - |
dc.subject | radiochromic film dosimetry | en |
dc.subject | intravascular brachytherapy | en |
dc.subject | radiation-therapy | en |
dc.subject | source trains | en |
dc.subject | quality-assurance | en |
dc.subject | calibration | en |
dc.subject | cancer | en |
dc.title | Patient and personnel dosimetry in endovascular radiotherapy with Sr-90/Y-90 sources | en |
heal.type | journalArticle | - |
heal.type.en | Journal article | en |
heal.type.el | Άρθρο Περιοδικού | el |
heal.identifier.primary | Doi 10.1093/Rpd/Nch496 | - |
heal.identifier.secondary | <Go to ISI>://000230595300007 | - |
heal.identifier.secondary | http://rpd.oxfordjournals.org/content/114/4/514.full.pdf | - |
heal.language | en | - |
heal.access | campus | - |
heal.recordProvider | Πανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών Υγείας. Τμήμα Ιατρικής | el |
heal.publicationDate | 2005 | - |
heal.abstract | fEndovascular brachytherapy (EVBT) is an established treatment to reduce the probability of restenosis after a percutaneous coronary intervention. The purpose of this study was to assess (1) the manufacturer's stated dosimetric data for Sr-90/Y-90 source trains to be used in EVBT and (2) the procedure-related radiation burden. The radiation fields in water around six Sr-90/Y-90 source trains were studied using phantoms made of 'solid water' and MD55-2 radiochromic films. The water equivalence of the phantom material was tested by applying quantitative computed tomography. Thermoluminescence dosemeters were used to assess personal radiation burden and crosscheck the dose distribution along the source trains. Technical failure was observed in one source train and this train was excluded from analysis. The measured dose rate in water at 2 mm radial distance was on average 8% higher than the manufacture's stated value (range of measured to stated values 1.05-1.15). The dose rate decreased exponentially with radial distance between 2 and 6 mm. The dose rate in contact with the source viewing window of the delivery devices ranged between 0.5 and 7.5 mGy h(-1). Low-energy photons were the main contributors to personal dose. | en |
heal.journalName | Radiation Protection Dosimetry | en |
heal.journalType | peer-reviewed | - |
heal.fullTextAvailability | TRUE | - |
Appears in Collections: | Άρθρα σε επιστημονικά περιοδικά ( Ανοικτά) - ΙΑΤ |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
Kalef-Ezra-2005-Patient and personne.pdf | 360.6 kB | Adobe PDF | View/Open Request a copy |
This item is licensed under a Creative Commons License