Please use this identifier to cite or link to this item: https://olympias.lib.uoi.gr/jspui/handle/123456789/16366
Full metadata record
DC FieldValueLanguage
dc.contributor.authorRoyle, G. J.en
dc.contributor.authorPapanestis, A.en
dc.contributor.authorSpeller, R. D.en
dc.contributor.authorHall, G.en
dc.contributor.authorIles, G.en
dc.contributor.authorRaymond, M.en
dc.contributor.authorCorrin, E.en
dc.contributor.authorVan der Stelt, P. F.en
dc.contributor.authorManthos, N.en
dc.contributor.authorTriantis, F. A.en
dc.date.accessioned2015-11-24T18:30:23Z-
dc.date.available2015-11-24T18:30:23Z-
dc.identifier.issn0168-9002-
dc.identifier.urihttps://olympias.lib.uoi.gr/jspui/handle/123456789/16366-
dc.rightsDefault Licence-
dc.titleDevelopment of a 2D silicon strip detector system for mammographic imaging using particle physics technologyen
heal.typejournalArticle-
heal.type.enJournal articleen
heal.type.elΆρθρο Περιοδικούel
heal.identifier.primary10.1016/s0168-9002(02)01566-8-
heal.identifier.secondary<Go to ISI>://WOS:000179230000006-
heal.identifier.secondaryhttp://ac.els-cdn.com/S0168900202015668/1-s2.0-S0168900202015668-main.pdf?_tid=050c5f614e29136b6dc46dbb85da6105&acdnat=1334229900_beca3761b71458cf5adc448dc6ec04e7-
heal.accesscampus-
heal.recordProviderΠανεπιστήμιο Ιωαννίνων. Σχολή Επιστημών και Τεχνολογιών. Τμήμα Βιολογικών Εφαρμογών και Τεχνολογιώνel
heal.publicationDate2002-
heal.abstract2D silicon strip sensors using particle physics readout technology have been evaluated as mammographic detectors. Two different versions of the APV series of front-end electronics were used that provided different noise levels. The sensors were evaluated using a typical mammography X-ray spectrum. The spatial resolution was evaluated using line pair test patterns and the modulation transfer function (MTF) was measured using the Edge Response Function. Low contrast performance was measured using the TOR(MAX) test object. Limiting spatial resolution of 52 mum was obtained and an MTF value of 0.1 at 16 1p/mm. The low contrast performance was estimated from 250, 500 pm and 6 min diameter objects and was found to be 11.5%, 7% and better than 3.8%, respectively. (C) 2002 Elsevier Science B.V. All rights reserved.en
heal.journalNameNuclear Instruments & Methods in Physics Research Section a-Accelerators Spectrometers Detectors and Associated Equipmenten
heal.journalTypepeer reviewed-
heal.fullTextAvailabilityTRUE-
Appears in Collections:Άρθρα σε επιστημονικά περιοδικά ( Ανοικτά)

Files in This Item:
File Description SizeFormat 
Royle-2002-Development of a 2D.pdf694.9 kBAdobe PDFView/Open    Request a copy


This item is licensed under a Creative Commons License Creative Commons